7. Replacing parts
ORTHOPANTOMOGAPH OP 2D
101
3. Remove the upper shelf top cover (4 x M4
screws) from the upper shelf.
3. Remove the stiffening plate (10 screws) from
the top of the upper self and then disconnect
all the cables that connect the rotating unit to
the L1600 - Z-motor board. Also disconnect
the Ethernet cable and the ground leads.
4.
CAUTION:
HEAVY OBJECT; THE FOLLOW-
ING TASK REQUIRES TWO PEOPLE.
Place some soft material on the ground so that
the rotating unit will not be damaged when it is
removed from the upper shelf and placed on
the ground.