2-4
4.
Span Cal.
Pressing this soft key (F5) just to the right of the Span
Cal option will set the Span of the pressure sensors. The software
assumes 5 psi is applied to the pressure sensors. Note: To ensure
pressure has been applied the software looks for an A to D count of
455 to 485 as displayed in the Service Mode in the left numerics
beside P1 & P2.
2.1.2 Battery Status Screen.
The Battery Status is brought up by pressing the F2 soft key (second
key down on left side of screen) from the SERVICE MODE screen.
Figure 2-3.
2.1.2.1 Battery Status
The Battery Status screen (see Figure 2-3) displays the State
of Charge, Terminal Voltage, each of the 4 Cell voltages,
current in milliamps, cycle count, and battery serial number.