2-8
Token
Description
Format
Event
14
Stop infusion button
pressed
#.# mL
Store VI
15
Cancel infusion button
pressed
#.# mL
Store VI
17
Alert Detected
Alert number. See
table 4
18
Hold time exceeded alarm
22
Pump door open has just
been opened
23
Pump Malfunction
Error number
See error number
Table 2.
29
Time or date was changed
Time when set time
screen was exited
39
History data corrupted
40
On Battery, or on AC Power
Switch to AC power or
Start/Switch on
Battery power.
AC = AC Power
B = Battery Power
49
Audible alarm silenced
Alarm Code
See Table 3.
50
Alarm Detected
Alarm Code.
See Table 3.
67
Power Off -with power
switch
68
Power Off - without power
switch (AC power or
Battery failure)
79
Pressure Unstable
81
Program Lost - re-enter
settings
83
Service mode activated
Service screen has been entered
87
Dose rate program –
Weight (0.1 kg – 999.9 kg)
#.# kg
When started from the DRC screen.
88
Dose rate program – Dose
(0.001 – 999.900)
0.001 mcg/kg/min, or
0.001 mg/kg/min, or
0.001 mg/kg/hr
When started from the DRC screen.
Store the appropriate units with the
dose value.
89
Dose rate program –
Concentration (0 – 99.990)
0.001 mg
When started from the DRC screen.
90
Dose rate program – Dilute
(1- 999)
1 mL
When started from the DRC screen.
91
Dose rate program – Rate
0.1 ml/hr
When started from the DRC screen.
92
Dose rate program – VTBI
0.1 ml
When started from the DRC screen.
93
Dose rate program – Drug
Drug index. See table
5.
When started from the DRC screen.
94
Dose rate program – Time
99 min.
When started from the DRC screen.
95
Recheck Settings
0 = Primary
1 = Secondary
2 = Bolus
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...