3-53
“v” firmware version. Press (F6) soft key; “Load Defaults”, note Com-
mand, Sent, and Rec. setting will scroll while defaults are loaded. Once
Scrolling stops, press (F5) soft key “Store” and verify Rec.: and Sent:
display “m”. Press the CANCEL key and set Keylock Enable; insure that
Keylock is also enabled on 3860 Pump.
3.9.5 Remote Pump Control Verification (3865)
3.9.5.1
ENTER and UP/DN key verification; Load test IV Set in 3850
Pump. With Remote Rate set to 100 and VTBI set to 10.
3.9.5.2
Start Control and Green Run Led Verification; With Test IV
Set loaded in 3860+ Pump, Rate set to 100, VTBI set to 10, and test
Pump displaying “Ready To Start”. Press START/STOP Channel A on
Remote and verify 3860+ Pump starts pumping and Green run LED
flashes on Remote.
3.9.5.3
Alarm Led Verification; With Pump running as described
above, occlude inlet on Pump and verify the following:
1. Pump stops
2. Inlet Occluded flashes in the upper portion of the screen
3. Audio alarm sounds
4. Red bar alarm LEDs flash
3.9.5.4
ALARM SILENCE Key Verification; Press the Alarm Silence
key and verify the audio alarm sound is silenced.
3.9.5.5
CANCEL Key Verification; With Pump stopped and red bar
alarm LED flashing, press the CANCEL Key and verify Alarm LED stops
flashing and “Ready to Start” flashes at the bottom of the screen.
3.9.5.6
MENU Key and Alarm Sound Adjustment Verification; Press
the MENU Key, then press the Alarm Volume soft key, and using the
Up/Dn arrow keys, verify the alarm volume can be raised, lowered and
saved (by pressing the ENTER key) to the appropriate setting.
3.9.6 Remaining Soft Key Verification
With Key Lock enabled in the 3860+ Pump Service Menu, press the
MENU Key and verify the following:
3.9.6.1
Verify pressing (F1) soft key “Dose Rate Calculation” displays
the Dose Rate Menu (press CANCEL).
3.9.6.2
Verify pressing (F3) soft key “KVO Rate” displays the KVO
Rate menu (press CANCEL).
3.9.6.3
Verify pressing (F4) soft key “Occlusion Limit” displays the
Occlusion Limit Menu (press CANCEL).
3.9.6.4
Verify pressing (F5) sof key “Lock Keys” (twice) returns to
normal display and all keys except MENU and OFF (0) are locked. Press
menu and (F5) soft key to “Unlock Keys”.
3.9.6.5
Verify pressing (F6) soft key “NEXT MENU” displays the next
menu with (F3) soft key “Set Comm Channel” selection (press CAn-
CEL).
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...