3-48
Verify Ready to Start is flashing in lower right of the Pump
display. Press the START/STOP Channel A key and verify the
Pump begins running as indicated by the pulsing “>” in the
upper portion of the display and Green flashing LED in the top
of the Door.
Press Go on the Infutest 2000. Allow Pump to run for a
minimum of 2 minutes, then press Stop on the Infutester.
Verify Average Flow Rate is 999 mL +/- 3% and record.
3.8.3.5 500 ml Flow Rate Test
Select Cancel Infusion with the bottom function key on the
Pump. Set the Pump Rate to 500 mL/hr., & VTBI to 62 mL.
Press Start on the Infutest 2000 and rotate flush syringe stop
to Prime position and inject about 5 cc of water into the
channel over about 5 seconds. Return stop to Test position.
Verify Ready to Start is flashing in lower right of the Pump
display. Press the START/STOP Channel A key and verify the
Pump begins running as indicated by the pulsing “>” in the
upper portion of the display and Green flashing LED in the top
of the Door.
Press Go on the Infutest 2000. Allow Pump to run for a
minimum of 2 minutes, then press Stop on the Infutester.
Verify Average Flow Rate is 500 mL +/- 3% and record.
3.8.3.6 125 ml Flow Rate Test
Select Cancel Infusion with the bottom function key on the
Pump. Set the Pump Rate to 125 mL/hr., & VTBI >200mL.
Press Start on the Infutest 2000 and rotate flush syringe stop
to Prime position and inject about 5 cc of water into the
channel over about 5 seconds. Return stop to Test Position.
Verify Ready to Start is flashing in lower right of the Pump
display. Press the START/STOP Channel A key and verify the
Pump begins running as indicated by the pulsing ">" in the
upper portion of the display and Green flashing LED in the top
of the Door.
Press Go on the Infutest 2000. Allow Pump to run for a
minimum of 5 minutes, then press Stop on the Infutester.
Verify Average Flow Rate is 125 mL +/- 3% and record.
3.8.3.7 25 ml Flow Rate Test
Select Cancel Infusion with the bottom function key on the
Pump. Set the Pump Rate to 25 mL/hr., & VTBI >100 mL.
Press Start on the Infutest 2000 and rotate flush syringe stop
to Prime position and inject about 5cc of water into the
channel over about 5 seconds. Return stop to Test position.
Verify Ready to Start is flashing in lower right of the Pump
display. Press the START/STOP Channel A key and verify the
Pump begins running as indicated by the pulsing ">" in the
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...