3-54
3.9.7
BATTERY CHARGE and Burn-In Verification (3865)
Install a new uncharged Battery (1133) and power on the 3865 while
holding down the “CANCEL” key.
3.9.7.1
Press the (F2) soft key and verify the following Battery Sta-
tus: State of Charge (%) xxx [0 to 2] Term. Voltage (V) xx.xxx [15 to
16.8].
Cell 1 x.xxx [3.8 to 4.25]
Cell 2 x.xxx [3.8 to 4.25]
Cell 3 x.xxx [3.8 to 4.25]
Cell 4 x.xxx [3.8 to 4.25]
Current (ma) xxx [+20 or greater (Note for later reference)]
cycle count x [0]
S/N (Verify serial number matches serial number on battery)
3.9.7.2
Record “Start Burn-in” time and leave 3865 on in this mode
for 12 hours.
3.9.7.3
After 12 hours, record Burn-in stop time. Verify if Battery is
fully charged and Battery Term. Voltage (V) in Battery Status display is
16.84 +0.03/-0.