3-50
3.8.5.1
Turn both 3860+ Pump and 3865 Remote Display/Charger
on.
3.8.5.2
Press the Menu key on the Remote Display/Charger
3.8.5.3
Press the NEXT MENU soft key to bring up additional menu
options, then press the Set Comm Channel soft key.
3.8.5.4
From the Radio Channel menu, select the desired Channel 1
through 6 by pressing the soft key next to that channel and verify that
Channel is now highlighted.
3.8.5.5
Press the Menu key on the MRidium 3860+ MRI IV Pump.
3.8.5.6
Press the NEXT MENU soft key to bring up additional menu
options, then press the Set Comm Channel soft key.
3.8.5.7
From the Radio Channel menu, select the same Channel that
was selected for the Remote by pressing the soft key next to that chan-
nel and verify that Channel is now highlighted.
3.8.5.8
Add an ID marker on both the Remote and Pump with the
Channel selected.
3.8.5.9
Verify communication between Remote and Pump by pressing
the Menu key on one and the Cancel key on the other, and observing
both display match and change simultaneously.
NOTE:
Each pump ships from the factory with a default setting of 63 mW
(+18 dBm) and at this level will penetrate most MRI magnet screen rooms.
Some MRI magnet screen rooms are tight enough to require the maximum
power level setting of 250 mw (+24 dBm).
To increase the power level to maximum reference paragraph 2.1.3.3 and
change command “wp” from 01 to 02.
A simple test of radio link robustness can be conducted utilizing the ‘open
field’ method. The pump and remote and be separated by 100 feet and verify
proper remote control operation. Notice that while scrolling numeric values,
the numbers roll smoothly will little pause or delay and that the link lost alert
is not triggered. Should the above test fail, check the antenna and internal
antenna cable (AC28) for damage.
3.9
Power Supply verification (3865)
(Reference IVPAB013 assembly drawing in section 5.)
Additional Test Item needed: TF009 Battery Simulator Assembly and cables.
3.9.1 DC Input Verification (3865)
3.9.1.1
Connect Line Power to 3865, measure DC output from power
supply between J5 pins 1 (Gnd) & 2 (+V) on the IVPAB013. Adjust DC
supply for 19V +/- 0.5V at IVPAB013, J5-2.
3.9.1.2
Press Keypad On (I) button. Verify and record the following:
IVP-AB013 TP 1 is 5V +/- 75mV. IVP-AB013 TP 2 for 3.3V +/- 75mV
3.9.1.3
Press Keypad Off (0) button for 2 Seconds.
Summary of Contents for MRidium 3860