2-7
The following table describes this history log output.
Token
Description
Format
Event
1
Power On
SW Revision string
At power up
3
KVO Rate changed
#.# mL / hr
4
Occlusion Pressure Alarm
Limit (downstream)
changed
# PSI
5
Alarm Volume Limit
changed
# (no units)
7
Dose Error Reduction Soft
Limit Over-ride
Refer to Table 6
8
Dose Error Reduction Hard
Limit Over-ride
Refer to Table 6
10
Infusion Rate
#.# mL/hr, P or PT
#.# mL/hr, S or ST
#.# mL/hr, K or KT
#.# mL/hr, B or BT
Only when Primary (P), Secondary
(S), KVO mode (K) or a Bolus (B) for
channels A or B is started or after
titration (T) is confirmed
11
Volume to be infused
#.# mL, P or PT
#.# mL, S or ST
#.# mL, K or KT
#.# mL, B or BT
Only when Primary, Secondary, KVO
mode or a Bolus for channels A or B is
started or after titration is confirmed
12
Volume infused
#.# mL, P
#.# mL, S
#.# mL, K
#.# mL, B
#.# mL
Only when Primary, Secondary, KVO
mode or a Bolus for channels A or B is
started, after titration is confirmed, or
the VI is cleared.