2-14
2.1.3.5 Language Option
On the 3860+MRidium 3860+ pump, pressing the Language
soft key (F5) will enable the user to scroll through the various
languages available for use on the MRidium 3860+ screen
display. Once the selected language is set in the Service
Menu, exit the Service Mode. On next power-up of pump,
selected language will be displayed.
2.1.3.6 Next Menu
Pressing the Next Menu soft key (F6) will change the display
screen to the second Service Menu option page (See Figue 2-
9).
Figure 2-8.
2.1.3.7 Flow Switch On
For MRidium™ 3860+ Pump’s, Flow Switch On should be
displayed as the first item on the screen.
2.1.3.8 Drug Table On
For MRidium 3860+ pumps, the Dose Rate Caculation table
can be turned ON or OFF by pressing the soft key next to
Drug Table (see Figure 2-8).
2.1.3.9 Edit Custom Drugs
For MRidium 3860+ pumps, the Edit Custom Drugs menu
allows user to program custom drug settings in the Dose Rate
Caculator Set-up Screen (see Section 1.6 for instructions on
programming custom drug settings).