3-9
3.2.4.11.3.2
Occlude inlet of IV Set above Pump and
verify Red Alarm LEDs are flashing and speaker
Alarm sounds on both Pump and Remote. Press
ALARM SILENCE key on remote, verify Alarm
sound is silenced, and Alarm LEDs stop flashing
on both Pump and Remote. Press Start Channel
A Remote and verify green (run) LED is flashing
on both Pump and Remote.
3.2.4.11.3.3
On Remote, press MENU key. Press (F1)
soft key to select Dose Rate Calculation and
verify Dose Rate A screen is displayed on both
Pump and Remote. Press CANCEL key.
3.2.4.11.3.4
On Remote press the (F3) soft key to
select Alarm Volume. Verify Alarm Volume
screen is displayed on Remote and “Adjusting
the Remotes’ Alarm Volume” is displayed on the
Pump. Verify pressing the Remotes’ Down arrow
key lowers the volume and causes the displayed
bar to move to the left of the screen. Pressing
Remotes’ Up arrow key increases the volume
and causes the displayed bar to move to the
right of the screen. Return displayed bar to the
center of screen, then press the ENTER key.
3.2.4.11.3.5
On Pump, press the (F3) soft key to
select Alarm Volume. Verify Alarm Volume
screen is displayed on Pump and Adjusting the
Pumps’ Alarm Volume is displayed on the
Remote. Pressing Pumps’ Up arrow key
increases the volume and causes the displayed
bar to move to the right of the screen. Verify
pressing the Pumps’ Down arrow key lowers the
volume and causes the displayed bar to move to
the left of the screen and at lowest alarm volume
is still audible. Return displayed bar to the center
of screen, then press the ENTER key.
3.2.4.11.3.6
Press (F6) Next Menu then press the
(F2) soft key to select Occlusion Limit. Verify
Occlusion Limit screen is displayed. Verify that
pressing the Down arrow key decrements 0.1 psi
with each press to 1 psi. Verify pressing the Up
arrow key increments 0.1 psi with each press to
10 psi and press Enter key.
3.2.4.11.3.7
Press the (F6) soft key to select Next
Menu. Press the (F1) soft key to select KVO rate.
Verify pressing the Up and Down arrow keys
selects from 1 to 5 mL/hr for the 3850 and from
0 to 5mL/hr for the 3860. Return setting to 1
mL/hr., then press the ENTER key.
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...