3-17
3.2.10.4 10 Key Pad Verification (3865 Remote) only.
3.2.10.4.1
On the 3865 Remote using 10 key pad with Rate
highlighted enter 123, verify tone with each key
pres
s
and then press enter verifying 123 entered in
rate.
3.2.10.4.2
On the 3865 Remote using 10 key pad with VTBI
highlighted enter 456, verify tone with each key
press and then press enter verifying 456 entered in
VTBI.
3.2.10.4.3
On the 3865 Remote using 10 key pad with Rate
highlighted enter 789, verify tone with each key
press and then press enter verifying 789 entered in
rate.
3.2.10.4.4
On the 3865 Remote using 10 key pad with VTBI
highlighted enter 900, verify tone with each key
press and then, press C and verify VTBI displays 0.
Enter 999 Using 10 key pad, then press Enter.
3.2.10.4.5
On the 3865 Remote using 10 key pad with Rate
highlighted enter 83.2, verify tone with each key
press and then press enter verifying 83.2 entered in
rate. Press Start A.
3.2.11 Battery Pack Maintenance
Over discharge, over charge and over current (short-circuit) are all
conditions detected by the internal “smart power gauge” of the battery.
These conditions all result in the battery shutdown (battery will not
deliver power) until placed into charge. Over discharge can cause cells
to outgas during subsequent charge cycles. Severe outgassing can
result in the battery pack swelling. Replace any battery pack with
visible signs of deformation or swelling.
It is important to understand the basic chemistry of lithium polymer
batteries. These batteries are a limited-life technology. The useful life
of the battery pack is affected by chemical processes which start the
day the each of the cells inside the battery pack are made. The life can
be further shortened by the way the cells are used and stored. These
cells will die faster if they are not held at some small level of charge
(50%) during non-use periods. If they are stored with any cell below
3.0V, a higher percentage of the cells will start to outgas and swell. If
they go to very low voltages, they may not charge well and make even
more gas.
The 1138 Operator’s Manual includes many cautions and suggestions
for caring for the battery. But lithium polymer batteries are sensitive
use and storage conditions, and inherently have a limited useful life.
From the chemical standpoint and normal distribution of failure, we
expect less than 5% should need replacement in less than 6 months. If
you see more than 10% of your batteries need replacement within 6
months of installation, we must look closely at the use and storage
conditions.
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...