3-24
Systems Test
3.5 Systems Test
Perform the following tests for a complete verification of the systems of the
mRidium IV pump family. If a major component is replaced, the test may be
resumed at point of failure, unless previous test steps are related with the
replaced component. IVPAB001, IVPAB002 in the 3860+ pump, IVPAB011
(3861 SideCar) and the AS100 pump assembly in both the 3860+ and 3861
are components that require the restart of a complete test. Note, the following
procedure is the same for both 3860+ Pump (Service Mode A) and 3861
SideCar (Service Mode B). When testing the 3861 SideCar, always remove all
power including 3860+ Pump battery when making connections for testing.
3.5.1 Reference Documents
Infutest 2000 Operation Manual
Sartorius Model GD603 or DG605 Scale Operation Manual
3860 Operation Manual REF 1138
3.5.2 Test Equipment Needed
(For complete detail system testing/calibration)
4.5 digit DMM (BK model 1670A) or equivalent.
Regulated DC Power Supply, 20+V/3A.
20+ MHZ Dual Channel Oscilloscope (Tectronix TDS210) or equivalent.
PC with serial port RS-232 capability.
Test Software provided on 1125 Service Manual CD.
Sartorius Model GD603 or GD605 Scale with RS-232 connection.
Infutest 2000 IV Pump Test set (or equivalent)
Iradimed 1000 Series IV Sets and IV pole.
IV Fluid Container (Distilled Water)
Closed Continuous Flow Circular/Looped Series 1000 IV Set.
Pressure gauge (0 to 10 psi) and pressure means ( ie: hand bulb).
IVPAB009 PCBA & Test Adapter Cable Set.
SideCar Pump Test Adapter Cable Set.
106 db RF Attenuator Cable.
Battery Test Cable.
TF009 Battery Simulator Test Fixture
Pronk Technologies OxSim, OX-1 SpO2 Finger Simulator (3860+ only)
Pull force gauge or spring balance
3.5.3 Basic Tools
#1 Phillips Screw Driver
#2 Phillips Screw Driver
PH00 Phillips
9/64” Allen Driver or T Bar
3 mm Allen Driver or T Bar
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...