3.2.6 Pump Flow Test.
Pump flow test can be preformed using a Infusion Pump Analyzer or a
graduated beaker. Place Pump on IV Pole with IVP1056 infusion set
connected to 1000 ML distilled water bag. Place output of IVP1056
infusion set into 100 ML empty graduated beaker or connect to Infusion
Pump Analyzer.
Load Primed infusion set (IVP1056) into Pump. Run the pump at a high
rate (1400 or 999 ML) insure there are no air bubbles traped in the Y
sites of the Infusion.
IF using graduated beaker be sure to empty after each flow test.
3.2.6.1
Set the Pump Rate to 1400 mL/hr., & VTBI >46.6 mL. Press
Start/Stop A on the Pump. Verify the pump is running as indicated by
the pulsing “>” in the upper portion of the display and Green flashing
LED in the top of the Door, and rate is set to 1400 mL/hr. Press Go in
the Infusion Analyser or monitor graduated beaker. The Pump will run
for approximately 2 min then go to KVO. Press Start/Stop A on the
Pump and verify average flow rate on infusion analyzer is 1400 ml +/-
5% or graduated beaker contains 46.6 ml +/- 5%. (44.27 - 48.93)
3.2.6.2
Set the Pump Rate to 999 mL/hr., & VTBI >33.3 mL. Press
Start/Stop A on the Pump. Verify the pump is running as indicated by
the pulsing “>” in the upper portion of the display and Green flashing
LED in the top of the Door, and rate is set to 999 mL/hr. Press Go in the
Infusion Analyser or monitor graduated beaker. The Pump will run for
approximately 2 min then go to KVO. Press Start/Stop A on the Pump
and verify average flow rate on infusion analyzer is 999 ml +/- 5% or
graduated beaker contains 33.3 ml +/- 5%. (31.634 - 34.965)
3.2.6.3
Set the Pump Rate to 500 mL/hr., & VTBI >16.6 mL. Press
Start/Stop A on the Pump. Verify the pump is running as indicated by
the pulsing “>” in the upper portion of the display and Green flashing
LED in the top of the Door, and rate is set to 500 mL/hr. Press Go in the
Infusion Analyser or monitor graduated beaker. The Pump will run for
approximately 2 min then go to KVO. Press Start/Stop A on the Pump
and verify average flow rate on infusion analyzer is 500 ml +/- 5% or
graduated beaker contains 16.6 ml +/- 5%.
(15.77 - 17.43)
3.2.6.4
Set the Pump Rate to 125 mL/hr., & VTBI >10.4 mL. Press
Start/Stop A on the Pump. Verify the pump is running as indicated by
the pulsing “>” in the upper portion of the display and Green flashing
LED in the top of the Door, and rate is set to 125 mL/hr. Press Go in the
Infusion Analyser or monitor graduated beaker. The Pump will run for
approximately 5 min then go to KVO. Press Start/Stop A on the Pump
and verify average flow rate on infusion analyzer is 125 ml +/- 5% or
graduated beaker contains 10.4 ml +/- 5%. (9.88 - 10.92)
3.2.6.5
Set the Pump Rate to 25 mL/hr., & VTBI >4.2 mL. Press
Start/Stop A on the Pump. Verify the pump is running as indicated by
the pulsing “>” in the upper portion of the display and Green flashing
LED in the top of the Door, and rate is set to 25 mL/hr. Press Go in the
Infusion Analyser or monitor graduated beaker. The Pump will run for
approximately 5 min then go to KVO. Press Start/Stop A on the Pump
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...