3-28
3.6.6.4 Zero Verification after Adjustment :
Open pressure gauge hand bulb and verify P1 and P2 now read 0 +/-
0.2. Repeat Tare Offset and Pressure Adjust until P1 and P2 reads dis-
plays 5 +/- 0.5 with 5 PSI applied and are 0 +/-0.5 with hand bulb
open (0 pressure applied).
3.6.7 Battery Verification :
Insure battery is easly removed and does not show signs of swelling.
(See Section 3.2.7.4 for swelled battery removal)
Press the F2 key And verify the following Battery Status, with charger
cable connected. Note : Pryor to test connect charger cable to pump
and allow battery to charge to at least 50%.
State of Charge (%) xxx [50 to 100]
Term. Voltage (V) xx.xxx [15 to 16.9]
Cell 1x.xxx [3.8 to 4.25]
Cell 2x.xxx [3.8 to 4.25]
Cell 3x.xxx [3.8 to 4.25]
Cell 4x.xxx [3.8 to 4.25]
(Note: All cells should be within .2 of each other)
Current (ma) xxx [+ 20 or greater]
cycle count x [ one or more]
S/N [Verify serial number matches 1133 Battery Pack serial Label
installed in the pump]
Disconnect 1122 Charger Cable and verify battery current (ma) is
approximately -240.
If battery shows signs of swelling or cells are not within 200 mv of each
other the battery should be replaced.
Press Cancel key to return To Service Mode A.
3.6.8 Bubble Detect Verification :
With no IV Test Set installed verify Bubble detect indicated by Upper
case “B” in the lower center of the display.
3.6.9 Verify No Bubble Detect :
Load a IVP1056 set filled with fluid and verify no Bubble detect by lower
case “b” indication in the lower center of the display with the IV Test
Set installed.
Summary of Contents for MRidium 3860
Page 2: ......
Page 4: ...1 4...
Page 8: ...iv...
Page 34: ...1 26...
Page 50: ...2 16...
Page 71: ...3 21 Figure 3 2...
Page 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Page 95: ...3 45 Return the Stop Cock to the Test Position...
Page 106: ...4 2 Figure 4 1...