IfU TECOTHERM NEO TN300 EN - 01.doc
Page 44 of 76
Ref No. TN300 EN-01
Preliminary temperature adjustment of the mattress / wrap , see Section 7.1
Prior starting treatment you may thermalize the system to a more appropriate
temperature. You may input a temperature different from 33,5°C by pushing key
Select.
33,5°C is taken as a standard set mattress temperature.
Note
After a longer device down time the fluid temperature is at ambient
temperature level. Starting operation the system will achieve a set mattress
temperature after some warm up or cool down time.
Adjust all necessary treatment parameters for the selected treatment mode.
Then push key
START
for starting Treatment Section I “Thermal preparation of the
mattress”.
Servo controlled complete treatment mode
Mattress temperature at start of treatment
Select
Apply standard settings or highlight, select and change settings
(Results in a re-warming of about 0.50 °C/hour)
Cancel
Start
33.5 °C
Duration of re-warming phase
7 h
Rectal temperature after re-warming phase
37.0 °C
Duration of Cooling Phase
72 h
Rectal temperature during Cooling Phase
33.5 °C
as required:
Note
: Treatment section “Themal preparation of Mattress” is optional
If set mattress temperature is reached system may stay ready for operation.
When neonate is placed and all is prepared push key
START
once again to start the
servo controlled treatment.
Example
How to perform hypothermia treatment in Treatment Mode 3
Exemplary approach: Treatment Mode 3 Constant Mattress Temperature