IfU TECOTHERM NEO TN300 EN - 01.doc
Page 62 of 76
Ref No. TN300 EN-01
Case 1
There is no indication of patient rectal temperature, in the
DIAGRAM feature
Measure
Check whether Rectal Temperature Probe is plugged correctly to the
socket
R.
If not plug connector tightly.
If cause is eliminated, rectal temperature reappears in the information box. Observe
display until measured rectal temperature approaches the set rectal temperature.
If no success probe may have a sensor break or is electrically short circuited.
In this event, replace rectal probe with a prescribed rectal temperature probe correctly
in the rectum of the patient, secure it. Plug the probe into socket
R
. Observe
temperature indicator at the display.
If no temperature indication appears repeat procedure with a new rectal probe.
If this fails you have to turn off the device, see Section 8.4.
Contact your service and inform.
Measure
Check whether Rectal Temperature Probe is plugged correctly to the
socket
R.
If not plug connector tightly.
If the cause is eliminated, rectal temperature reappears in the information box. Observe
display until measured rectal temperature approaches the set rectal temperature.
If no success probe may have a sensor break or is electrically short circuited.
In this event, replace rectal probe with a prescribed rectal temperature probe correctly
in the rectum of the patient, secure it. Plug the probe into socket
R
. Observe
temperature indicator at the display.
If no temperature indication appears repeat procedure with a new rectal probe.
If this fails you have to turn off the device, see Section 8.4.
Contact your service and inform.
Case 2
All temperatures are indicated. Rectal temperature deviates more than
0,5°C from the set rectal temperature.
Example
Indicated rectal temperature is lower than
set rectal temperature 33,5°C.
Note
In treatment sections III Cooling phase and IV Re- warming phase
a deviation of +/- 0,5°C is allowed (alarm limit).
Measure
Check whether rectal probe is placed in its correct position or slipped out
completely or partially. The more it is slipped out the lower is the detected
and indicated temperature (approaching ambient temperature.)
If slipped out place probe into correct position and secure it.
------
30,5