IfU TECOTHERM NEO TN300 EN - 01.doc
Page 73 of 76
Ref No. TN300 EN-01
If necessary
Turn Off TECOTHERM NEO device, see Section 8.4
If necessary
Put TECOTHERM NEO device onto a level table or support.
Take empty fill-up set, connect its QDC couplings to the QDC ports (6) at the device
front face. Turn bottle until cap is directed upwards. Lower fill up bottle in a position
lower than the TECOTHERM NEO device. Keep this position.
Start operation of the TECOTHERM NEO device according this IfU. Select treatment
mode 3 in the MENU. Fluid will be pumped into the refill set bottle. Continue pumping
until optical and audible signals “Fluid level too low, Refill” appear and no fluid flows
into the bottle.
Stop device operation see Section 8.4, turn off.
Unplug fill-up set. Waste dispose fluid into a drain.
Fill bottle with fresh TECOmed fluid. Close bottle cap tightly.
Start refilling according to section 8.5 of this IfU.
Then repeat emptying procedure as described above.
Final step: Refill TECOTHERM NEO as described in section 8.5 of this IfU.
Substitution of TECOmed fluid is finished, TECOTHERM NEO is ready for operation.
13.4 Substitution of TECOmed fluid in mattresses and cool wraps
Mattresses and Cool Wraps completely or partially filled and stored for a longer time
should be subject of TECOmed fluid substitution every 2- 3 months.
Detais see Section 8.6.
13.5
Software Update
Ensure that the TECOTHERM NEO applies latest software versions.
Customers will be informed on software updates by the manufacturer and/ or
authorised service provider.
13.6
Check/ calibration of temperature probes
To ensure proper operation and inteded use of the TECOTHERM NEO system use
only probes authorised by the manufacturer TEC COM GmbH.
Rectal and Skin Temperature Probes are delivered as autoclavable applied parts.
We recommend checking and calibration of the reusable temperature probes within
every 2 years, in accordance with Technical Specification / Manual of the temperature
probe manufacturer.