11.2 Temperature alarm
Temperature alarm
The display changes from Temperature indication LARGE NUMBER to DIAGRAM
feature indicating the alarm symbol. Audible and optical alarms appear.
Servo controlled complete treatment mode
Details
Set Point 33.5
32.7
Set Temperature
33.5
Elapsed time of cooling phase 02:34 of 72 hours
23.03.2010
08:55:42
Options
Mattress Temperature
Rectal Temperature
Skin Temperature
32.7
36.8
-----
Servo controlled complete treatment mode
15
20
25
30
35
1
2
3
4
5
6
7
8
9
hour
15
20
25
30
35
Current Rectal Temperature
Temperature alarm, Alarm paused
Caused by
Rectal Temperature probe disconnected from the device
Rectal Temperature probe is detached from patient
System operation temperature deviates more than
0,5°C from set
rectal emperature
Mattress temperature incorrectly measured
Fan cooling insufficient
Power of Central Cooling Module insufficient.
NOTE
When alarm is appearing push key T5 to silence audible
alarm (AUDIO paused,
is appearing at the display.)
There are now
8 minutes
to resolve the problem.
Elimination/ Management
Attention
Check immediately
whether Rectal Temperature probe is correctly
connected to the device and correctly positioned and secured in the
patient.
The operator should check and analyse the temperature profilies in the DIAGRAMM
feature. Look at first the temperatures in the upper part of the display and check
whether indicated temperatures are corresponding to the treatment section temperature
settings and are plausible and make sense.