IfU TECOTHERM NEO TN300 EN - 01.doc
Page 41 of 76
Ref No. TN300 EN-01
Next Step
Connect mattress / cool wrap to the TECOTHERM NEO
Prior to connecting check whether
mattress / wrap is filled completely. If empty or only partially filled, see Section
8.6 for filling instructions.
defects like punctures and flaws are existing. If fluid escapes replace or repair
defect mattress/ wrap.
tubing of hoses and mattress’ / wrap is kinked or situation may appear to kink.
Connect QDC couplings of hoses to the QDC counterparts of the mattress. Then plug
to the ports (6) at the device right lower front part, see figure. If they do not engage
properly push metallic unlocking keys at the QDC, then repeat plug procedure.
Figure
Connection to ports (6)
Note
To unlock push metallic unlocking keys downwards.
Place completely filled mattress/ cool wrap onto a
10 - 20 mm thick plastic- foam
material that has good thermal insulation.
Note
In case of Incubators see Section 6.
Then place a thin protective interlayer with plastics coating at the bottom downwards
onto the mattress. Protective interlayer should be somewhat larger than the mattress
and fully cover the mattress with a border for wrapping the mattress.
Unlocking Keys