IfU TECOTHERM NEO TN300 EN - 01.doc
Page 23 of 76
Ref No. TN300 EN-01
All temperatures / time dates are recorded/ logged and can be read out/ transferred to
an USB stick, see section 7.3 for the USB port.
NOTE
All parameters can be changed from set position at ANY TIME should the need
arise. Changes will be stored on the TECOTHERM NEO and can be seen on later
analysis.
Attention
Fallback Case
If treatment is interrupted or any reason for stopping operation in Treatment Mode 2
SERVO Controlled Constant Rectal Temperatue you may continue using Treatment
Mode 3 Constant Mattress Temperature which is a non-servo controlled mode.
Note
Such treatment is a true alternative solution to continue treatment under
Treatment Mode 3 conditions (Fallback Case)
Note
Study details of treatment mode 3 prior to change treatment, see below.
Note
Transition from Automatic Mode to Treatment Mode 3 Constant Mattress
Temperature must be performed by the operator:
TECOTHERM NEO rectal temperature probe *) should be disconnected and
removed. It is inactive
BCT monitoring requires installation of another independent accurate calibrated
temperature probe (rectal probe or bladder catheter sensor) with a separate display
showing the rectal temperature.
Note
Operator may continue to use the skin sensor connected to port “
S
” of the
TECOTHERM NEO device as an auxiliary means if no other probes are available.
That temperature is displayed on the screen in the small parameter boxes in the
upper part of the display feature DIAGRAM and as temperature profile in the display
feature DIAGRAM, see section 7.3 of this IfU.
Change to the Main MENU and select treatment mode 3 Constant Mattress
Temperature Mode.
If transition to fallback mode must be done during treatment section II (cooling
down): To cool down fast and efficiently to reach BCT 33,5°C select low mattress
temperatures (12 – 15°C). Then follow the instructions of treatment mode 3, see
next part below.
If transition to fallback mode must be done during treatment phase III or IV: Follow
instructions of treatment mode 3, see this section below.
*)
If no independent rectal temperature probe is available you may use
TECOTHERM NEO rectal probe. That temperature is displayed at the screen
in the
small parameter box
, upper part of the display feature DIAGRAM.
Numbers are small!