IfU TECOTHERM NEO TN300 EN - 01.doc
Page 60 of 76
Ref No. TN300 EN-01
SF
System Alarm
Alarm feature 2
SF
LED (5) in the lower part of the front panel is activated
Display screen is shut off / not illuminated (dark)
Caused by
Communication error or problem between Operational Board and
subordinated Control Board
The Control Board then creates an acoustic alarm, and SF LED (5) is lit.
System continues operating with the set parameters unless operator is
silencing audible alarm pushing key T5
AUDIO paused
The Control Board then 10 seconds later is
switching off
the device
The operator can now analyse the situation. Turn
ON
the
TECOTHERM NEO device to continue treatment operation.
If SF indicator alarm reappears switch system
Off.
After 5 seconds
Turn On device again and try to continue in Treatment Mode 3 Constant
Mattress Temperature as the Fallback Mode, see section 7.1.
If no success Shut
Off
the system and contact your service partner.