IfU TECOTHERM NEO TN300 EN - 01.doc
Page 39 of 76
Ref No. TN300 EN-01
Important entries, instructions, notes and failures are displayed as turquoise-
highlighted Pop- Up Windows
TECOTHERM NEO
A check of the internal functions has been performed and did not
Now the alarm functions will be checked. Can you see the various
Symbols and clearly hear the two alarm sounds?
detect any malfunction.
No
Yes
ERROR
It is not allowed to run the device without working alarm
Confirm
functions. It will shut off in 30 seconds (press "Confirm" to shut
off immediately).
Important conditions and preconditions to perform treatments are displayed on the
screen. The operator has to
Confirm
or
Cancel.
NOTE: A rectal probe is required to be inserted prior to treatment.
Servo controlled complete treatment mode
Confirm
Cancel
For reference, an additional skin temperature probe is strongly
Please confirm the requirement for a rectal probe by pressing the
Please ensure its correct placement prior to starting treatment.
recommended.
"Confirm" button!
Treatment Data Selection and Transfer/ USB socket
To read out and export treatment data and data log files select Entry
Export
Treatment Data
in the Main MENU. To perform export plug an USB stick into the USB
port (9) at the rear side of the device. Then follow MENU instructions.
The operator gets an information on the screen whether data export was successful.
After successful export unplug USB stick.
For service
: Socket (9) is an USB interface port. Plugging USB hardware stick is for
updating firmware and detailed data export.