Hydraulic Module and Circulation System
The Hydraulic Module is made of a pump, pressure valve, inner fluid container, quick
disconnect couplings to connect hoses and mattresses, cool wraps, and an internal
flow meter to detect circulation flow volume. Pump is driving the TECOmed fluid to
circulate through the mattresses/ applied parts. Operation pump pressure is limited to
0,5 bar.
Micro Computer controlled Operating Board and Control Board, MENU
TECOTHERM NEO is furnished with
an Operating Board
and a
MENU
as user
interface, and with a
Control Board
to control its operation features. Both boards are
based on their separate powerful micro computers which are in permanent
communication (master OP – slave CB.)
TECOTHERM NEO has three main operation and treatment modes. The operator can
select operation modes using the MENU.
A large display serves as
user interface
for the operator. MENU operations and
treatment modes are visualized on the display screen.
The user either selects, confirms or modifies treatment modes, treatment options,
operations and settings using MENU operation
Arrow Keys
to move to MENU entries.
Pushbuttons below display screen enable performing instructions like Select, Confirm,
Cancel, Apply, Start etc.
Current instructions and entries are highlighted turquoise.
Display with MENU
Arrow Cursor
Arrow keys