IfU TECOTHERM NEO TN300 EN - 01.doc
Page 65 of 76
Ref No. TN300 EN-01
Case 3
Volume of circulating fluid slightly too little or too little
Lack of fluid is not so big.
Low Fluid Level
alarm is not activated. Try to fill up
approx. 100 ml
fluid to enhance circulation (fill up until greater air bubbles disappear.)
Proceed as follows (see for details Section 8.5 and 8.6, too) :
Connect only
one connector QDC of the fill-up bot
tle to the
right port
of the
refill port QDC counterparts (7) at the device front face. Turn bottle until cap is
directed downwards. Lift fill up bottle in a position ½ m higher than the
TECOTHERM device. Keep this position.
Observe whether symbol
No Flow
disappears.
Note
: 5 sec
after disappearance of
No Flow symbol
Disconnect bottle
connector QDC from right port (7) ). Lower fill-up bottle and turn it back into its
normal vertical position. Unscrew bottle cap (1/2 revolution) for pressure
compensation (a few seconds). Then close bottle cap tightly.
If you do not succeed
If possible: Replace TECOTHERM NEO device by a replacement unit
or a spare unit. Start treatment according IfU.
Otherwise stop treatment and immediately contact your local service for assistance.