IfU TECOTHERM NEO TN300 EN - 01.doc
Page 67 of 76
Ref No. TN300 EN-01
Case 2
Check / inspection of mattresses and hoses
Inspect the mattress, hoses, couplings for traces and indications of liquid/ moisture.
If there is a big leakage, replace this defective component immediately.
If you do not find a leakage to define the lack of fluid proceed as follows:
Refill TECOmed liquid acc. Sections 8.5 and 8.6 .
Symbol
must vanish. Observe the display screen.
If leakage is small you may replace the defective component. If no spare
components are available, temporarily resolve the problem to finish patient treatment:
Close the small leak by means of an adhesive tape or appropriate plasters
(non permeable.) After finishing the treatment, you have to replace the defective
part or repair the mattress using the repair kit. Mattress must be empty and dry.
Note
If the
alarm repeatedly appears
, check mattress, hoses, connectors and
tubing once again to detect if and where fluid escapes permanently. In such cases, see
instructions here in this Section 11.4.
Attention
After refilling, the TECOTHERM NEO is ready for use.
Usually the added fluid has a different temperature than
the system fluid.
Hence: Temperature alarm as section 11.2 may appear.
We recommed to push key T5 to silence audible alarm (AUDIO paused,
is
appearing at the display.)
Within a few minutes the system will reach the set operation temperature.