Indicating Temperatures at the Display screen
Running the 3 main treatment modes, see section 7.2, treatment temperature is shown
on the screen in a large size (display feature LARGE SIZE NUMBER)
Rectal temperature
is displayed when selecting treatment mode 1 or 2.
Mattress
temperature
is shown when selecting treatment mode 3.
Treatment modes 1 and 2
As long as measured Rectal Temperature deviates
more than 0,5 °C from the Set Point the rectal temperature is appearing
RED
.So for
example if Set Point is 33,5°C a measured rectal temperature of 36,4 °C is displayed
RED
.
Only when deviation is < 0,5°C colour changes to
GREEN
.
Servo controlled complete treatment mode
Details
Set Point 33.5
36.4
Servo controlled complete treatment mode
Details
Set Point 33.5
33.4
Treatment mode 3
As long as measured Mattress Temperature deviates more
than 0,5 °C from the Set Point the mattress temperature is appearing
RED
.So for
example if Set Point is 30,0°C a measured temperature of 34,1 °C is displayed red.
Only when deviation is < 0,5°C colour changes from
RED
to
BLUE
.
Constant Mattress Temperature Mode
Details
Set Point 30.0
34.1
Constant Mattress Temperature Mode
Details
Set Point 30.0
30.2
60 seconds after starting treatment in each of the main
treatment
modes the display
feature
DIAGRAM
is changed to the feature
LARGE SIZE NUMBER
.
Here shown for treatment mode 2 Servo Control Constant rectal temperature
.
Servo control mode (constant rectal temperature)
Details
Set Point 33.5
33.6
Set Temperature
Servo control mode (constant rectal temperature)
33.5
Constant rectal temperature for 01:34 hours
23.03.2010
08:55:42
Options
Mattress Temperature
Rectal Temperature
Skin Temperature
33.6
35.4
-----
15
20
25
30
35
1
2
3
4
5
6
7
8
9
hour
15
20
25
30
35