IfU TECOTHERM NEO TN300 EN - 01.doc
Page 59 of 76
Ref No. TN300 EN-01
TECOTHERM NEO
Confirm to continue with "Yes"
No
Yes
Power failure during treatment - do you wish to continue with
or
Start a new treatment with "No".
previous treatment?
If System alarm
SF
remains or is appearing again after some time
Shut off device by unplugging Mains cable
Inspect the system according following instructions
SF
System Alarm
Symbol is displayed at the screen, display feature
DIAGRAM,
accompanied by optical and audible alarms.
Caused by
Failure in the TECOTHERM NEO system
- Defective Central Cooling Module
- Defective pump
- Temperature exceeding limits
- Internal communication problems between Operational System and
Control system
Elimination/ Management
Step1 did not solve the problem !
Shut off device by unplugging Mains cable
Or
Push key T1 Options, turn to the Main MENU. Select Power Off by
pushing key T6.
NOTE
It is not allowed to
Turn On
the device
Contact your local service provider, for assistance.
TECOTHERM NEO Main Menu
Ser.-Nr. 2010/08/01
Rev. 023/01.32
Highlight and Select Function Required:
Servo controlled complete treatment mode
Servo control mode (constant rectal temperature)
Constant Mattress Temperature Mode
Alarm check
Export treatment data
Service
Power Off
Select
Language
If possible Replace TECOTHERM NEO device by a replacement unit or a spare unit.
Put it into operation according IfU.