IfU TECOTHERM NEO TN300 EN - 01.doc
Page 46 of 76
Ref No. TN300 EN-01
Treatment is started
Normally TECOTHERM NEO reaches the set rectal temperature 33,5°C after 25 – 35
min (small mattress 30x 45 cm) and 30 – 35 min (cool wrap. Have in mind that cooling
down rate is depending on weight of the patient and on the ambient conditions/
temperature: the higher the more time is required.
TECOTHERM NEO is progammed to reach target rectal temperature 33,5°C or any
other selected temperature as soon as possible if no other rate input is chosen by the
operator (time to reach final temperature > 0 h ).
The operator may observe the temperature – time profile in the display feature
DIAGRAM. Or temperature values in the small parameter boxes in the upper part.
Independent on treatment mode feature DIAGRAM changes after 60 seconds
automatically to feature LARGE SIZE NUMBER.
Treatment is running
NOTE:
We recommend that during treatment operator should
from time to time monitor patient temperatures.
This is especially important during transition from cooling section III to
re warming section IV.
Treatment is executing according to and depending on selected treatment mode and
applied treatment parameters. All treatment data like temperatures / time dates are
automatically recorded/ logged and can be read out/ transferred to an USB stick.
NOTE: All parameters can be changed from set positions pushing key T1 Options and
Select, at ANY TIME ,should the need arise.
The display shows the entry Change the current settings., see below. You may go back
to the original settings pushing key T3 Back . Changes will be stored on the
TECOTHERM NEO and can be seen on later analysis
Options during operation
Changes the current settings
Alarm check
Select
The current operation continues to run.
Highlight and select the desired option:
Immediately start the re-warming phase
End treatment and save data
Back