IfU TECOTHERM NEO TN300 EN - 01.doc
Page 57 of 76
Ref No. TN300 EN-01
After Switch On of the TECOTHERM NEO the operation readiness of the alarm system
is automatically checked through self- testing.
Display screen is showing the Information
TECOTHERM NEO
A check of the internal functions has been performed and did not
Now the alarm functions will be checked. Can you see the various
Symbols and clearly hear the two alarm sounds?
No
Yes
detect any malfunction.
SF
Confirm YES when you see the various symbols and clearly hear the two alarm sounds
If not confirm NO. The display shows a Pop Up window ERROR It is not allowed to
run the device without working alarm functions.
TECOTHERM NEO will shut off at after 30 seconds.
Push key T1 Confirm to immediately shut off the device !
TECOTHERM NEO
A check of the internal functions has been performed and did not
Now the alarm functions will be checked. Can you see the various
Symbols and clearly hear the two alarm sounds?
detect any malfunction.
No
Yes
ERROR
It is not allowed to run the device without working alarm
Confirm
functions. It will shut off in 30 seconds (press "Confirm" to shut
off immediately).
Operator can select at any time during treatment operation an Alarm Check using
MENU. Use arrow keys to highlight this function.
Options during operation
Changes the current settings
Alarm check
Select
The current operation continues to run.
Highlight and select the desired option:
Immediately start the re-warming phase
End treatment and save data
Back