Function
State:
Description
Off (open)
open:
always open
Ready for operation
closed:
Emission switched on, calibration process
inactive, no error
open:
Emission switched off or calibration process
active or error
Emission on
closed:
Emission switched on
open:
Emission switched off
Error
closed:
Error
open:
No error
Error or warning
closed:
Error or warning
open:
No error or warning
Run-up
closed:
Run up
open:
no run-up
Calibration request
closed:
No external calibration: Calibration required
When external calibration or "CAL check":
Request "Open or close external calibrated
leak"
open:
No request: Calibration not required
Calibration active
closed:
Device is to be calibrated.
open:
Device is not to be calibrated.
Calibration stable
closed:
Calibration completed with calibration leak
open:
Assignment not stable or calibration is
inactive
Cathode 2
closed:
Cathode 2 is active
open:
Cathode 1 is active
Measuring
closed:
Measuring (ZERO is possible, all setpoint
outputs switch depending on the leak rate.)
open:
Standby or emission disabled (ZERO is not
possible, all setpoint outputs will return
"Leak rate threshold value exceeded".)
Open
open:
always open
Sniff
closed:
Operation mode Sniffing active
open:
Operation mode Vacuum active
Setpoint 1
closed:
Measured leak rate exceeds the setpoint 1
open:
Measured leak rate undercuts the setpoint 1
Setpoint 2
closed:
Measured leak rate exceeds the setpoint 2
open:
Measured leak rate undercuts the setpoint 2
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......