6 | Operation
INFICON
88 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
ð
After completion there is an automatic restart of the system.
6.10.3 Update the software in expert mode
ü
Supervisor rights
1
> Update > Update operating unit/basic unit > Expert
ð
Software versions already available on the device are shown in a list.
2
Highlight any of the software versions as required and proceed to step 5.
ð
Resetting to an older software version is possible.
3
If alternatively you want to add a new software version, connect a FAT32
formatted USB flash drive with the update file to one of the USB ports of the
device.
4
In order to load the new software version, press on
.
5
To activate the new version, select the desired list entry and press
.
ð
After completion there is an automatic restart of the system.
6.11 Use external monitor
You are able to use an external touch monitor instead of the internal monitor. You can
use either the internal or the external monitor.
ü
You have a control unit with HDMI interface, for the illustration see “Control unit
ü
You have a touch monitor with connections for HDMI and USB, furthermore an
appropriate power supply. HDMI is required to transmit the images, USB is required
to transmit the sound.
1
Connect the HDMI cable and the USB cable of the touch monitor with the
connector block of your leak detector, see also “Connections for accessories
33]“. For proper operation, make sure that the cable
length does not exceed 3 m.
2
> Display > Switch monitor
3
Press the "External Monitor" button.
ð
The picture will be displayed on the external monitor.
Alternative procedure
► Instead of changing the monitor display, you can connect the external monitor
before switching on the device.
ð
Then the picture will be displayed on the external monitor without changing
settings.
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......