7 | Warning and error messages
INFICON
108 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
Type
Notification
Possible sources of
error
Remedy
W710
MSB temperature is too high The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located
Air filter dirty
• Clean the air vents or replace the filter
plates
Fan blocked or
defective
• Contact customer service
E711
Max. MSB temperature
exceeded
The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located
Air filter dirty
• Clean the air vents or replace the filter
plates
Fan blocked or
defective
• Contact customer service
W901
Maintenance: TMP bearing/
lubricant
Maintenance interval
for TMP bearing/
lubricant exceeded
• Contact customer service
W903
Maintenance: Calibration
leak expired
Maintenance interval
for calibration leak
exceeded
• Contact customer service
W904
Maintenance: Sniffer tip filter Maintenance interval
for sniffer tip filter
exceeded
• Contact customer service
W910
Maintenance: Backing pump Maintenance interval
for backing pump
exceeded
• Contact customer service
W920
Maintenance: Exhaust filter
Maintenance interval
for exhaust filter
exceeded
• Contact customer service
W925
Maintenance: Air filter
Maintenance interval
for air filter exceeded
• Contact customer service
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......