6.4 Measure
ü
The inlet flange on the upper side of the device is prepared for the pending
measurement. Your test specimen or the SL200 sniffer line is normally connected
there.
ü
Alternatively, the sniffer line SL3000 is connected to the back of the device
(optional). In this case, the inlet flange on the top of the device should be closed
with a blank flange.
ü
Possible alternatives to the operating possibilities on the device are set up
(optional):
I/O module, bus module, remote control, wireless LAN connection to the notebook
or tablet. Also see "Accessories and spare parts [
detector via web browser (wireless LAN) [
133]".
1
Switch on the leak detector via the power supply switch.
2
Make sure that the desired operation mode is set (see "Select operation mode
3
Make sure that the correct basic settings and the settings for the current
measurement are carried out. See "Basic settings" and "Settings for the
measurements".
4
When performing a calibration, note the 20 minute warm-up time. See
5
In standby mode, initiate the measurement process by pressing the START
button.
ð
If you want to ensure that a current internal background is used in the
following measuring operation, please observe the notes in chapter “Change
72]“.
6
Keep track of the measurement results either as line, bar or circle chart, see
"Presentation of the measurement screen [
55]" and "Assembly of the
touchscreen".
7
To display the maximum leak rate (Q
max
) on the measurement screen during a
measurement, press the START button again. Also see "START button [
25] "
8
To measure small leak rates more easily, press the ZERO button when
required, see "Set and use the function ZERO [
69]".
9
Record the measured values if necessary, see "Switch on data recording and
78]" and "Design of touchscreen [
10
Switch the instrument off.
See also
2
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......