The modes "Standard" or the mode "Inverse" are available for selection.
To mode "Inverse": In the following table overview, the arrow pointing the
transfer direction in the column "Transition" must be reversed.
5
Save
.
Function
Transition
Description
VENT
inactive→ active:
Vent inlet (only in standby state)
CAL
inactive→ active:
active→ inactive:
Start calibration.
Apply value for background and finish calibration.
No function
-
The input is not functioning. Recommended setting of unused inputs.
Clear
inactive→ active:
Erase warning or error message / cancel calibration.
Sniff
inactive→ active:
active→ inactive:
Enable sniffer mode.
Enable vacuum mode.
Purge
inactive→ active:
active→ inactive:
Switch-on purge (only in standby state)
Switch-off purge (only in standby state)
Start
inactive→ active:
Change to the measurement mode.
Start/Stop
inactive→ active:
active→ inactive:
Change to the measurement mode.
Switch to Standby.
Stop
inactive→ active:
Switch to Standby.
ZERO
inactive→ active:
active→ inactive:
Switch ZERO on.
Switch ZERO off.
10.2.4 Configuring digital outputs
The digital outputs PLC-OUT 1 ... The available functions can be assigned in any way
necessary to the 8 I/O module.
ü
Supervisor rights
1
> Setup > Accessories > I/O module > Digital outputs configuration
2
Select the desired "Digital output".
ð
The outputs PLC-OUT 1 - 8 are available.
3
Select the desired "Function", see the table below.
4
Select the desired "Mode".
ð
The modes "Standard" or the mode "Inverse" are available for selection.
To mode "Inverse": In the table overview the description of the state "open"
must be replaced with the description of the state "closed".
5
Save
.
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......