7 | Warning and error messages
INFICON
100 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
Type
Notification
Possible sources of
error
Remedy
E340
Emission error
Pressure too high with
aged cathode
• Contact customer service
W342
Cathodes not connected
Both cathodes
defective during self-
test
• Contact customer service
Plug of ion source not
connected
• Contact customer service
W350
Suppressor not connected
Suppressor cable not
connected or defective
• Contact customer service
E352
Preamplifier not connected
Preamplifier defective
or cable not connected
• Contact customer service
W358
Alternate preamplifier
ranges
Preamplifier defective
• Contact customer service
Unusual leak rate
waveform
• Contact customer service
W359
Overdriven preamplifier
Ion source or
preamplifier defective
• Contact customer service
Mass spectrometer
contaminated
• Contact customer service
W360
Preamplifier output signal
too low
Ion source defective
• Contact customer service
Mass spectrometer
contaminated
• Contact customer service
W361
The offset voltage of the
preamplifier is too high
Preamplifier defective
• Contact customer service
W362
Preamplifier range error
Preamplifier defective
• Contact customer service
MSB defective
• Contact customer service
W390
500 GOhm amplifier resistor
out of range
Preamplifier defective
• Contact customer service
MSB defective
• Contact customer service
E400
TMP error message
Unhandled error
message from the TMP
• Contact customer service
W401
TMP warning message
Unhandled warning
message from the TMP
• Contact customer service
E402
No communication with
turbo pump converter
TMP cable defective or
not connected
• Contact customer service
TMP converter
defective
• Contact customer service
MSB defective
• Contact customer service
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......