6.2.12 Activating or deactivating calibration requests
If the option "calibration request” is not active (factory setting), you only get a
calibration request in the following cases:
• A SL3000 sniffer line with a new serial number has been connected.
• The operation mode has been changed via the operating unit and there was no
calibration in the new operation mode within the last 24 hours.
If the option "calibration request" is active, you will get a calibration request
subsequently in the following cases:
• The period of the device is longer than 30 minutes and the temperature deviation
compared to the last calibration is larger than 5 Kelvin.
• There was no calibration within the last 24 hours in the currently selected
operation mode.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Miscellaneous
2
To switch on further calibration requests activate the option “Calibration
request“. To switch off deactivate this option.
3
Save
.
6.2.13 Extending or reducing operation options
You can always operate the device via touchscreen. Furthermore, you can change the
factory setting, which enables the operation via buttons on the control panel and via
the remote control.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Miscellaneous
2
Under “Local control“ you make your choice in the selection field between:
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......