INFICON
Operation | 6
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
57 / 146
6.2.6.3 Changing the presentation of the circle graph
Fig. 12: Presentation of the circle chart
ü
Operator or Supervisor rights
1
> Display > Circle graph
2
If you want to fix the number of displayed decades in the circle chart, under
"Decades“ select the number of displayed decades. Only effective when
automatic scaling is disabled.
3
If you want to dynamically adjust the display range to the measured leak rate,
activate automatic scaling via the option “Autoscale“.
4
In the “Lower diagram limit“ field, select the desired lower limit for the circle
chart. This limit is also effective if the “Autoscale“ option is activated.
5
Save
.
6.2.6.4 Change general display settings
ü
Operator or Supervisor rights
1
> Display > General
2
If very small leak rates are of no interest to your application and you want to
raise the lower leak rate limit, select between 1 - 6 decades in the “Raise lower
leak rate limit”. Raising the lower leak rate limit can facilitate the assessment of
the leak rate display.
3
To set whether to display the internal background leak rate in the “Standby”
window, enable or disable the “Display background at standby“ option.
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......
Page 146: ......