6 | Operation
INFICON
86 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
, to activate a set of parameters.
, to delete a set of parameters.
6.9.3 Exporting or importing sets of parameters
You are able to transfer saved sets of parameters from the internal memory to an
attached USB flash drive and to import the sets again.
ü
Supervisor rights
1
> Sets of parameters > Manage sets of parameters
2
To export all sets of parameters to a USB flash drive (formatted in FAT32) if
needed, press
below the list.
3
To import all sets of parameters from a USB flash drive into the device, press
below the list.
ð
All sets of parameters on the USB flash drive will be saved in the device, if
the number of 10 is not exceeded. Using a PC you can delete redundant sets
of parameters on the USB flash drive before transferring.
ð
You must not modify parameters of a set of parameters on the USB flash
drive before importing. You cannot transfer these files to the device after the
modification.
6.10 Updating the software
The device has two different software versions: One for the user interface and one for
the basic unit. Each has its own independent versions number.
Summary of Contents for 550-520
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Page 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Page 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Page 145: ......