4-4
Remote Control
Setting up a Measurement
Setting up a Measurement
To set up the measurement system, compile the interface code with the Visual
Basic files and the HP supplied
TESServer.dll
file. The
TESServer.dll
file allows your
software to read and write test parameters to and from the test parameter
database file,
testparm.mdb
, without requiring knowledge of the database’s
underlying structure.
C A U T I O N
The test parameter database file,
testparm.mdb,
is a complicated structure.
Damage to the file can render the HP 86037B unusable until the database is
replaced or repaired. For this reason, direct access of the
testparm.mdb
database
is not recommended or supported by HP.
Communication between the
TESServer.dll
on your computer and the databases
on the system is via a direct network connection to the system’s hard drive.
You’ll need to supply this network connection.
Setting up a measurement requires the following three steps:
1
From
testparm.mdb, e
xtract the parameter collection that will be passed to the
test.
2
Edit the test parameter values in the collection.
3
Save the collection to
testparm.mdb
after all editing is complete.
When the measurement is executed, the collection is retrieved by the test
executive from the test parameter database and used to perform the measure-
ment.
Extract the test parameters
The
TESServer.dll
can be installed on your remote computer in two ways. A com-
plete set of files can be installed using the installation file on the accompany-
ing CD-ROM. Or, you can copy the
TESServer.dll
file along with its companion
files
tes.ini, io.mdb,
and
security.mdb
from the system
C:\HPCD
directory. Once
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...