7-11
Reference
Measurement Repeatability
Measurement Repeatability
This section compares the measurement repeatability of HP 86037B with and
without the multi-wavelength meter. The measurements were performed on
the same spool of 22-km dispersion-shifted fiber. The wavelength range was
from 1525 nm to 1575 nm in 1 qnm steps. The data were analyzed with the
five-term Sellmeier fit.
The standard deviation for
λ
0
is actually higher for the measurements taken
with the multi-wavelength meter than for those without. This is most likely
because the measurements made with the multi-wavelength meter take
longer, and therefore are more susceptible to fiber length drifts. The particular
HP tunable laser source used in the measurement was so accurate that it did
not introduce much wavelength error. These results are based on one system,
therefore we cannot guarantee them as specifications for all systems since the
absolute wavelength accuracy of HP’s tunable laser sources is specified to be
±0.1 nm. For best accuracy, a system with a multi-wavelength meter is still
recommended.
Table 7-2. Repeatability Data for 20 Runs With the Multi-Wavelength Meter
Parameter
λλλλ
0
(nm)
S
0
[ps/ (nm
2
××××
km)]
Mean 1550.8938
0.0733825
Maximum 1550.9062
0.0734366
Minimum 1550.8829
0.0733237
Max-min 0.0233 0.0001129
Standard Deviation
0.008116
0.0000318
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...