4-12
Remote Control
Setting up a Measurement
Save the test parameters
When editing is complete, any changes need to be written to the parameter
database. The subroutine
StoreLevelParms
requires that the first parameter be a
collection and the second be an integer corresponding to the level in the test
plan hierarchy. Since the parameter collection for the Test level is a TEScol-
lection, rather than a standard Visual Basic collection, you must apply the
“Items” method or use the method:
TESMngr.ParmMngr.LoadTestParms
as done in
the example file in order to extract the underlying collection from within the
TEScollection.
Table 4-1. Integer Parameter Definitions for StoreLevelParms
Test Plan Hierarchy Level
Integer
System level
3
Test Plan
2
Test Group
1
Test level
0
Example code for writing database parameters
Public Sub save_parm_values()
'
' Purpose: Returns parameters to the database.
'
With TESMngr.ParmMngr
With TESMngr.Controller
.ActiveTestPlan = sActiveTestPlan
.ActiveTestGroup = sActiveTestGroup
End With
.StoreLevelParms colSystemParms, 3
.StoreLevelParms colTestPlanParms, 2
.StoreLevelParms colTestGroupParms, 1
.StoreLevelParms colTestParms, 0
End With
End Sub
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...