5-11
Maintenance
Verification Tests
3
After the
Preset
is complete, the following parameters appear on the network
analyzer’s display:
CH1, S11, log MAG, 10dB, REF 0dB, START 30kHz, STOP 3GHz
(or
6GHz for Option 006).
If the self test fails, refer to the HP 8753E Network Analyzer service manual
for instructions on testing the individual components of the network analyzer,
or call your HP service center.
To run the the network analyzer operator’s check
The operator’s check determines that the source is phase locked across the
entire frequency range and that all three samplers are functioning properly.
Figure 5-3. Operator’s Check for the HP 8753 Network Analyzer
1
Disconnect all devices, peripherals, and accessories (including adapters and
limiters) from the HP 8753 network analyzer.
2
Switch on the network analyzer.
3
Press the
Preset
button on the network analyzer.
4
Press the
System
button on the network analyzer.
5
On the network analyzer:
a
Press the softkey corresponding to the
Service
menu.
b
Press the softkey corresponding to the
Tests
menu.
c
Enter,
21
, on the network analyzer keypad.
d
Press the, x1, button on the network analyzer.
e
Press the softkey corresponding to the
External Tests
function.
6
Press the
Execute Test
key on the network analyzer.
7
At the prompt, connect the equipment as shown in
, with power to
Table 5-5. Equipment Required
Equipment
Model Number
20 dB attenuator
HP 8491A Option 020
RF cable set
HP 11851B
Two-way power schooner
HP 11667A Option 001
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...