7-12
Reference
Measurement Repeatability
Table 7-3. Repeatability Data for 20 Runs Without the Multi-Wavelength Meter
Parameter
λλλλ
0
(nm)
S
0
[ps/ (nm
2
××××
km)]
Mean 1550.8850
0.0733461
Maximum 1550.8906
0.0734181
Minimum 1550.8789
0.0732212
Max-min 0.0117
0.0001969
Standard Deviation
0.00354
0.0000553
Table 7-4. Comparison Between the Mean Values
Parameter
Mean
λ
λ
λ
λ
0
(nm)
Mean S
0
[ps/ (nm2
××××
km)]
With wavelength meter
1550.8850
0.0733461
Without wavelength meter
1550.8938
0.0733875
∆
0.0088
0.0000364
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...