6-6
Installation
Installation
Step 2. Set line voltage selector switch
This diagram shows cabling on an Option 012 dual-band system with Options
113 and 116. For option 114, the wavemeter is removed.
1
Make sure that the system is unplugged from the main power supply.
2
Confirm that the line voltage selector switch on the HP 83427A test set is set
to the voltage of the mains supply. The line voltage selector switch is located
next to the rear-panel line cord plug. Use a flat blade screwdriver to open the
voltage selector door and change the voltage selection tumbler. For 115V
operation, use type N 1.0A/250V fuse. For 230V operation, use a 0.75A fuse. A
special fuse holder is provided for the 0.75A fuse.
W A R N I N G
HP 83427A: The power cord is the disconnecting device. It disconnects
the mains circuits from the mains supply before other parts of the
instrument. Alternately, an externally installed switch or circuit
breaker (which is readily identifiable and is easily reached by the
operator) may be used as a disconnecting device. The front panel
switch is only a standby switch and is not a LINE switch
(disconnecting device).
C A U T I O N
HP 83427A: Always use the three-prong AC power cord supplied with this
product. Failure to ensure adequate earth grounding by not using this cord may
cause product damage.
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...