5-7
Maintenance
Troubleshooting Common Problems
Otherwise, perform the procedure
“To check the test set” on page 5-7
.
To check the test set
W A R N I N G
Prevent any accidental exposure to the laser beam. Prior to
disconnecting any optical fibers on the TLS, ensure that the TLS is
turned off.
To accurately troubleshoot the system, make sure the TLS is working properly
prior to testing the test set.
Refer to “To check the TLS” on page 5-6
.
1
Use a microscope to check the connections on the test set.
Connections for Accurate Measurements” on page 5-20
for information on
recognizing damaged connectors.
2
Run the self test for the HP 8753E Network Analyzer.
network analyzer self test” on page 5-10
. If the self test passes, troubleshoot the
test set. If the test fails, the network analyzer needs repair
To check the computer
❒
For problems with the personal computer that are unrelated to the rest of the
HP 86037B, refer to the computer manual that was shipped with this system,
or call your local IT department.
❒
Troubleshoot the computer’s connection to the HP 86037B:
a
Make sure that the BNC cable (or a serial cable on the dual band system) is
connected to the test set properly.
b
Ensure that the GPIB and computer addresses are correct. The correct
GPIB and I/O addresses for the computer and system are located in
.
❒
Perform the following steps:
a
Disconnect the BNC cable (or a serial cable on the dual band system) from
the test set.
b
With a voltage meter, measure the voltage from the cable. It should be read-
ing between 2 and 4 volts.
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...