1-11
Quick Start
Measuring Grating Devices
Measuring Grating Devices
In this section, you’ll learn how to make measurements on grating based
devices such as Bragg fiber gratings. It is recommended that you do not skip
this section. Since the HP 86037B, by default, is set for making measurements
on fiber devices, before measuring grating devices, perform the following
steps:
1
Select
Absolute Group Delay
Measurement
in the
Grating Related
Parameters
area in the main
window. (By default, relative
group delay is measured.)
to “Absolute Group Delay Measurement” on page 2-14
.
In addition to selecting absolute group delay, this selection turns off the
review data window’s curve fit selection. Although, curve fitting is a standard
method used to reduce the effects of noise when measuring fiber based
devices, there is currently no standard for grating based devices. You can, of
course, export the data and apply your own corrections.
2
Set the
Test Modulation Frequency
to the
Manual
setting.
to configure this
setting.
For details on determining the proper RF test modulation frequency,
“Setting the RF modulation frequency” on page 1-12
.
3
Use larger wavelength steps during test setup normalization compared to
during the measurement. This reduces overall test time.
Since patch cords used during normalization have very low dispersion, the
system is able to interpolate accurate values for use with grating measure-
ments. For example, if you are measuring a grating with 5 nm wavelength
range, you could use 1 nm steps during normalization and 1 pm steps during
measurements.
4
Expect large jumps in measured dispersion.
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...