3-11
Normalization
Adjusting the Power Monitor
5
Enter 0.0 in the
Power constant
field. Click
OK
.
6
Press the
Local
key on the TLS (tuneable laser source) to put it in local
operation mode.
7
On the TLS, set the wavelength to 1558.5 nm and the power to 0 dBm.
8
Connect a short fiber-optic cable from the
SOURCE OUTPUT
connector to an
optical power meter. Press the
Active
key on the tuneable laser source.
9
Measure the optical power at 1558.5 nm on the power meter and record the
reading P
1
.
P
1
: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ dBm
10
Disconnect the fiber-optic cable from the power meter and connect it to the
RECEIVER INPUT
port on the HP 83427A test set.
11
Click the
Power Monitor
button on the main window and record the power
reading P
2
.
P
2
: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ dBm
12
Calculate the difference using the following equation:
∆
P = P
1
– P
2
P
∆
: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ dBm
For example:
∆
P = (–12 dBm) – (–8 dBm)
∆
P = –4 dBm
13
Close the
Power Monitor
window.
14
Choose
Power Monitor Settings
from the
Service
menu.
15
In the
Power Monitor Setup
dialog box, enter the following values:
Power constant:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ∆
P (as calculated above)
Under range limit: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –43 dBm
Insertion loss reference: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –15 dBm
TLS power:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 dBm
TLS wavelength: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1558.5 nm
16
Click
OK
.
17
Click the
Power Monitor
button on the main window. Verify that the displayed
power is within ±0.5 dB of value P
1
.
18
Click the
Set zero
button from within the
Power Monitor
window.
19
Close the
Power Monitor
window. The TLS automatically reverts to system
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...