2-36
Operation
Checking Test Fiber Stability
Checking Test Fiber Stability
Changes in environmental temperature can affect the length of the fiber-optic
cables that you are testing. Because changes in length degrade measurement
results, use this procedure to determine if the fiber-optic cable has stabilized
and is ready to test. For information on the theory behind this test, refer to the
sidebar
“Measurement theory” on page 2-38
.
1
Connect a patch cord “through” cable between the
SOURCE OUTPUT
and
RECEIVER
INPUT
.
2
Click
Modulator Bias
and then click
OK
.
3
When the modulator bias adjustment is complete, remove the thru fiber and
connect the fiber to be tested between the test ports.
4
On the front panel of the HP 8168 tunable laser source, press the
LOCAL
key.
When you complete this procedure and return to using the mouse or key-
board, control of the instruments return to the HP 86037B. The network ana-
lyzer and tunable laser source are automatically initialized for all of the system
measurements.
5
Verify the laser output power is set to at least –4 dBm at a wavelength in the
passband of the fiber-optic cable that you are testing.
6
On the front panel of the HP 8753 network analyzer, press the
LOCAL
key.
7
Enter the CW frequency by pressing:
MENU, CW FREQ,
and
1 G/n.
8
Enter the number of points by pressing:
MENU, NUMBER of POINTS,
and then
401
X1.
9
Enter a convenient sweep time, such as 60 seconds, by pressing:
MENU, SWEEP
TIME,
and then
60 X1.
10
Press
MENU, TRIGGER MENU,
and
CONTINUOUS.
11
Press
MENU
and then
MEASURE RESTART.
12
Wait until the first sweep is complete, and then press:
SCALE/REF
and then
AUTOSCALE.
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...