3-2
Normalization
Normalization
Normalization
Normalization removes any systematic effects of the test setup from the mea-
surements. The normalization path must include any fiber-optic patch cables
used to connect the device being tested to the front-panel
SOURCE OUTPUT
and
RECEIVER INPUT
connectors. As
shows, there are two different normal-
ization procedures and two adjustments that are necessary.
The laser normalization and power monitor adjustments should be performed
when the system is installed. They do not need to be performed before each
individual measurement. You should perform the test setup normalization
once a day, unless the system has been turned off since the last test setup nor-
malization. The modulator bias adjust can be automatically configured to run
each time a test setup normalization is performed.
Table 3-1. Recommended Interval between Normalizations and Adjustments
Procedure
Recommended Interval
Test setup normalization
Once a day or before any critical measurement.
“Normalizing the Test Setup” on page 3-4
.
Modulator bias adjustment
Before any measurement session. You can select to have this
adjustment procedure run automatically whenever a device
normalization is performed.
Refer to “Adjusting the Modulator
.
Laser normalization
When the system is installed or the tunable laser source is
replaced.
Refer to “Normalizing the Laser” on page 3-8
.
Power monitor adjustment
When system is installed or the tunable laser source is replaced.
Refer to “Adjusting the Power Monitor” on page 3-10
.
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...