13. Demodulator Adjustment
8. A 1
sine wave should be observed on the oscilloscope. Rotate the volume knob on the
front panel of the spectrum analyzer until the amplitude of the 1
signal is at about
150
(3 divisions on the oscilloscope).
9. Adjust
FM DEMOD for a maximum peak-to-peak response on the oscilloscope.
10. Press (LINE) to turn the spectrum analyzer off. Disconnect the test cable from
CONNECT HERE
Figure 2-22. Demodulator Adjustment Locations
Adjustment Procedures 2-55